IMAX 112
Specification
Fluoroscopic Capacity:
Max rated capacity
Tube Current 4mA
Tube Voltage 110kV
Automatic Fluoroscopy
Tube Voltage:40kV~110kV adjust automatically
Tube Current:0.3mA~4mA adjust automatically
Manual Fluoroscopy
Continuous tube Voltage:40kV~110kV
Continuous tube Current:0.3mA~4mA
Pulse Fluoroscopy
Continuous tube Voltage:40kV~110kV
Continuous tube Current:4.1mA~8mA
Photography Capacity:
- Max rated capacity - 3.5 KW
- Tube voltage, mAs - 40kV~110kV
- 20~63mA
- 1~125mAs
Plateholder size - 200mm X 250mm or 250mm X 300mm
X-ray Tube:
X-ray tube special for High frequency
- Fixed anode x-ray tube with 2 focus: Large focus: 1.5, small focus: 0.6
- Inverter Frequency: 40kHz Thermal capacity: 30kJ (40kHU)
Video System:
- Image Intensifier Image Intensifier made by TOSHIBA (91)
- CCD vidicon Imported CCD Vidicon with ultra-low luminosity
- Monitor Horizontal 1000 lines and vertical 800 lines
Bandwidth: 12.5MHz, Image/sec: 25
- CCU (central control) Recursive filter: K=8, 7 images storage, image upright, image overturn, positive & negative image; LIH (last image freeze), and OSD (monitor display)
Structure:
- Direction-wheel - +90o revolution, can change the moving direction of the unit
Ascending & descending range of pillar - >400mm
C-Arm - Forward and backward movement: 200mm
Revolution around horizontal axis: +1800
Revolution around vertical axis: +15 Slip on orbit: 120(+900~ -300)